Applied Photophysics Ltd

Chirascan Circular Dichroism Spectrometer

Chirascan Circular Dichroism Spectrometer Chirascan Key Features near-UV and Visible Performance

Example of near-UV visible spectra (1)

Circular dichroism spectrum of vitamin B12

Scan range   600 - 200 nm
Spectral bandwidth   1 nm
Step size   1 nm
Pathlength   1 mm
Concentration   1 mg/mL
Solvent   Water
Time per point   2 s
Total N2 flow rate   5 litres min-1

Example of near-UV visible spectra (2)

Circular dichroism spectrum of tris(ethylenediamine) cobalt chloride

Scan range   800 - 250 nm
Spectral bandwidth   1 nm
Step size   1 nm
Pathlength   10 mm
Concentration   1 mg/mL
Solvent   Water
Time per point   0.5 s
Total N2 flow rate   5 litres min-1

Example of near-UV visible spectra (3)

Circular dichroism spectrum of camphor sulphonic acid

Scan range   300 - 180 nm
Spectral bandwidth   0.5 nm
Step size   0.5 nm
Pathlength   1 mm
Concentration   0.6 mg/mL
Solvent   Water
Time per point   1.75 s
Total N2 flow rate   5 litres min-1

Example of near-UV visible spectra (4)

Circular dichroism spectrum of (R)-3-methylcyclopentanone

Scan range   350 - 175 nm
Spectral bandwidth   0.1 nm
Step size   0.1 nm
Pathlength   10 mm
Concentration   Vapour
Solvent   -
Time per point   1 s
Total N2 flow rate   5 litres min-1

 

Return to Chirascan product page.

Chirascan - the new standard in circular dichroism

Chirascan Circular Dichroism Spectrometer
Chirascan Circular Dichroism Spectrometer
Chirascan Circular Dichroism Spectrometer