Applied Photophysics Ltd
Manufacturer and supplier of circular dichroism, stopped-flow, and laser flash photolysis spectroscopy instruments.

SX20 Stopped-Flow Spectrometer - Site Requirements

SX20 Stopped-Flow Spectrometer Site Requirements for Installation

Below are the site requirements for Applied Photophysics SX-Series Stopped-Flow Spectrometers. These requirements must be fulfilled before the installation engineer arrives to install the spectrometer. If there are any questions regarding these requirements please contact the Customer Care Department.

 
     
     
  Get Adobe Reader
Download Adobe Acrobat Reader Here

Bench area

  • SX20 System: 2m x 0.7m
  • SX20 System with SK1E option: 2.5m x 0.7m

Electrical requirements

Total Current. Peak VA (minimum value of UPS)
220/240 Volt systems    1.6 A 700 VA
110 Volt Systems 3.2 A 700 VA

Gas requirements

Compressed gas supply, usually nitrogen or air, delivered to the system at between 8 and 9 bar. (124 to 140 psi). The system will be delivered with suitable, 1/4 inch ID, high pressure tubing to connect the regulator to the Sample Handling Unit.

This gas supply is used to drive the sample handling unit drive rams. In systems fitted with ozone producing xenon lamps, part of the gas supply is bled of to purge the lamp housing. In this instance the gas supply must be nitrogen.

Thermostating

If there is a requirement for temperature control of the sample handling unit then a circulating thermostat bath and tubing must be available. The tubing that connects to the sample handling unit should have an internal diameter of 8mm.

 

 

SX.20 Stopped-Flow Spectrometer
SX.20 Stopped-Flow Spectrometer
SX.20 Stopped-Flow Spectrometer